Calibrate and verify instrument
performance
■
Calibration and verification schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
■
Calibration descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
■
View the calibration status and set reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
■
Perform ROI/uniformity, background, and dye calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
■
Perform instrument verification using RNase P plates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
■
Calibrate custom dyes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
■
Calibrate for a custom melt curve experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Calibration and verification schedule
The instrument is factory-calibrated and does not require calibration at installation. To ensure optimal
performance, perform calibrations at the recommended frequency.
Note:
After instrument installation, we recommend performing instrument verification using the
provided RNase P plate.
IMPORTANT!
Perform calibrations and instrument runs under the environmental conditions that
are specified in “Environmental requirements”
. Exposure to extreme temperatures can
adversely affect the instrument performance and shorten the life span of the instrument components.
To set the calibration frequency for the instrument, touch
Settings
4
Maintenance and
Service
4
Calibrations
4
History and Reminders
4
Edit
4
Exp interval field
.
Calibration
Recommended frequency
ROI/Uniformity
• Every two years (recommended)
• Always perform new Background and Dye calibrations after an
ROI/Uniformity calibration.
Note:
Performing an ROI/Uniformity calibration invalidates all other calibrations.
5
52
QuantStudio
™
1 Plus Real-Time PCR System Installation, Use, and Maintenance Guide