3.
In the
Summary
tab:
• Click
Calibrations
to view the status of each calibration type.
•
(Optional)
Click
Calibration Reminders
to set the calibration reminder time table and enter
the notification email address(es).
Note:
More than one email address can receive the calibration reminders.
Note:
The settings are automatically saved.
•
(Optional)
In the
Downloads
section, click
Maintenance Summary.pdf
to download the
calibration status report.
4.
In the
Calibrations History
tab:
• View the history of each calibration type.
•
(Optional)
Click to download the calibration history report.
Perform ROI/uniformity, background, and dye calibrations
Workflow: Calibration
Note:
The ROI and uniformity calibrations use the same calibration plate.
Perform an ROI calibration
▼
Perform a uniformity calibration
You are automatically prompted to perform background calibration.
▼
Perform a background calibration
Perform any time that ROI calibrations are current.
▼
Perform system dye calibrations
Perform any time that ROI and background calibrations are current.
▼
(
Optional
) Perform custom dye calibrations
Perform any time that ROI and background calibrations are current.
Chapter 5
Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations
5
QuantStudio
™
1 Plus Real-Time PCR System Installation, Use, and Maintenance Guide
55