![Thermo Finnigan LTQ Hardware Manual Download Page 161](http://html.mh-extra.com/html/thermo/finnigan-ltq/finnigan-ltq_hardware-manual_1104223161.webp)
System Shutdown, Startup, and Reset
'JOOJHBO-52
________________________________________________________ Resetting the MS Detector
_______________________ Finnigan LTQ Hardware Manual ______________________
5-9
5IFSNP
&-&$530/$03103"5*0/
5.5 Resetting the MS Detector
If communication between the MS detector and data system computer is lost,
it may be necessary to reset the MS detector using the Reset button on the
power panel. When you press the Reset button (for 3 s), an interrupt in the
embedded computer is created. This causes the embedded computer to restart
in a known (default) state. See Figure 5-1 on page 5-2 for the location of the
Reset button.
The procedure given here assumes that the MS detector and data system
computer are both powered on and operational. If the MS detector, data
system computer, or both are off, refer to the topic Starting Up the System
after a Complete Shutdown on page 5-6.
To reset the MS detector, press the Reset button located on the power panel
for 3 s. Make sure the Communication LED is extinguished before releasing
the Reset button. When you press the Reset button (for 3 s), the following
occurs:
•
An interrupt on the embedded computer causes the CPU to reboot. All
LEDs on the front panel of the MS detector are off except the Power LED.
•
After several seconds, the Communication LED is illuminated yellow to
indicate that the data system and the MS detector are starting to establish
a communication link.
•
After several more seconds, the Communication LED is illuminated
green to indicate that the data system and the MS detector have
established a communication link. Software for the operation of the MS
detector is then transferred from the data system to the MS detector.
•
After 3 min the software transfer is complete. The System LED is
illuminated either green to indicate that the instrument is functional and
the high voltages are on, or yellow to indicate that the instrument is
functional, and it is in Standby.