MS Detector Maintenance
'JOOJHBO-52
_________________________________________________ Cleaning the Q0 and Q1 Ion Guides
_______________________ Finnigan LTQ Hardware Manual _____________________
4-37
5IFSNP
&-&$530/$03103"5*0/
Reinstalling the Top Cover of the MS
Detector
Reinstall the top cover of the MS detector as follows:
1. Open the front door of the MS detector by loosening the Allen screw on
the right side of the door with an Allen wrench or hex-head ball driver.
2. With one hand under the center of the top cover, place the top cover over
the MS detector such that the front of the cover is about 1.2 cm (0.5 in.) in
front of the front of the MS detector.
3. Slide the cover back until it is flush with the front doors (when they are
closed).
4. Tighten the two fasteners to secure the top cover to the chassis.
5. Close the front door of MS detector. Tighten the Allen screw on the right
side of the door with an Allen wrench or hex-head ball driver.
6. Reconnect any tubing between the syringe pump and the API source to
accommodate your instrument configuration.
Go on to the next topic: Starting Up the System.
Starting Up the System
Start up the system as described in the topic Starting Up the System After a
Complete Shutdown in the System Shutdown, Startup, and Reset chapter.