Daily Operation
Things to Do After Operating the LTQ
@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@'JOOJHBO-52
5IFSNP
&-&$530/$03103"5*0/
3-12
______________________ Finnigan LTQ Hardware Manual _______________________
b. In the LTQ Tune Plus window, choose
Setup > Inlet Direct
Control
(or click on the AS/LC direct control button). The Inlet
Direct Control view appears.
c.
Select the LC tab and click on
(or
Pump Off
or
Stop Pump)
to
stop the LC pump.
2. From the LTQ Tune Plus window, choose
Control > Standby
(or click
on the On/Standby button) to put the MS detector in Standby
.
Ensure that
a septum seals the entrance to the ion transfer capillary.
3. Open the gas ballast valve on the rotary-vane pump by turning it to
position
|
. Refer to the manual that came with the pump for the location of
the gas ballast valve.
4. Allow the pump to run for 30 min with the gas ballast valve open.
5. After 30 min, close the gas ballast valve by turning it to position O.
Go on to the next topic: Emptying the Solvent Waste Bottle.
Emptying the Solvent Waste Bottle
The solvent level in the solvent waste bottle should be checked on a daily
basis. If necessary, empty the solvent waste bottle. Dispose of the solvent
waste in accordance with local and federal regulations.
On Standby