Daily Operation
Things to Do After Operating the LTQ
@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@'JOOJHBO-52
5IFSNP
&-&$530/$03103"5*0/
3-8
_______________________ Finnigan LTQ Hardware Manual _______________________
9. After 5 min, turn off the API source by placing the MS detector in
Standby: From the LTQ Tune Plus window, choose
Control > Standby
(or click on the On/Standby button) to put the MS detector in Standby
.
Go on to the next topic: Placing the System in Standby Condition.
Placing the System in Standby Condition
Use the following procedure to place the LTQ system in the standby
condition:
1. From the LTQ Tune Plus window, choose
Control > Standby
(or click
on the On/Standby button) to put the MS detector in Standby
.
The System
LED on the front panel of the MS detector is illuminated yellow when the
system is in Standby.
2. Leave the MS detector power on.
3. Leave the LC power on.
4. Leave the autosampler power on.
5. Leave the data system power on.
Go on to the next topic: Flushing the Ion Sweep Cone and Ion Transfer
Capillary.
Flushing the Ion Sweep Cone and Ion
Transfer Capillary
You need to clean the ion sweep cone (or spray cone) and the ion transfer
capillary on a regular basis to prevent corrosion and to maintain optimum
performance of your API source. A good practice is to flush the ion sweep
cone and ion transfer capillary at the end of each operating day after you flush
the sample transfer line, sample tube, and API probe with a 50:50
methanol:water solution from the LC. (Refer to the topic Flushing the
Sample Transfer Line, Sample Tube, and API Probe on page 3-6.) If you
are operating the system with nonvolatile buffers in your solvent system or
high concentrations of sample, you might need to clean the ion sweep cone
and ion transfer capillary more often.
You do not need to vent the system to flush the ion sweep cone and ion
transfer capillary. To clean the ion sweep cone and the ion transfer capillary,
do the following:
1. Turn off the flow of liquid from the LC (or other sample introduction
device) to the API source. To turn off the flow of liquid from the LC to the
API source, do the following:
a.
Choose
Start > Programs > Xcalibur > LTQ Tune
to open the
LTQ Tune Plus window.
On Standby