4
Creating a Method
Creating a Method for the ISQ Mass Spectrometer
38
ISQ User Guide
Thermo Scientific
3. Set the
Ion Source Temperature
. You can enter a value between 0 and 350 °C. The
optimal temperature depends on the analyte. Higher temperatures will keep the ion
source cleaner, but will lead to increased fragmentation, which may reduce sensitivity. For
most compounds, a source temperature of 275 °C (default) is adequate.
4. Select the
Acquisition Threshold
checkbox and enter a value for the minimum peak
height for the data file, if needed. If your peak has an intensity that is below this
threshold, it will not be stored. This setting may help reduce noise, but it may also alter
the reported isotope ratios because the smaller isotope signals will be preferentially
reduced.
5. Select
EI
from the
Ionization Mode
pull-down menu.
6. In the Run Completion group, select the action you want to occur at the end of a run:
Note
For best results, tune the instrument at the same temperature you will run the
analyses in your method.
Note
Only use CI if you have installed a CI ion volume in the ISQ instrument and you
have connected CI reagent gas to your system. If you have CI, select a CI Gas Type from
the pull-down menu and set the CI Gas Flow. (There will be a two-minute delay when
you change ports for the CI Gas Type.) Typical values for Methane CI for NCI are
1.0–1.5 mL/min. For Methane CI for PCI, the gas flows are typically 1.5–2.0 mL/min.
Summary of Contents for ISQ
Page 1: ...ISQ User Guide 1R120555 0003 Revision D January 2013 ...
Page 22: ......
Page 54: ......
Page 82: ...4 Creating a Method Creating a Method for the GC 60 ISQ User Guide Thermo Scientific ...
Page 124: ......
Page 148: ......
Page 152: ......