7
Optimizing Your Method
Modifying an Automatic Tune
Thermo Scientific
ISQ User Guide
97
Figure 81.
Configuring the Tune Resolution
•
Mass
—Use this pull-down menu to select the ion to be used for tuning.
•
Peak Width
—Use this field to enter the target peak width.
•
Measure at %
—Use this pull-down menu to select the location on the peak at which
you want to measure the target peak width.
–
10
—Select this option to measure the width at 10% of the peak height.
–
50
—Select this option to measure the width at 50% of the peak height.
•
Lens Tune Relation
—Use this pull-down menu to set the occurrence of the
resolution tune parameters before or after a lens tune.
–
Before and After
—Select this option to use the same resolution parameters
before and after a lens tune.
–
Before
—Select this option to use the resolution parameters before a lens tune.
–
After
—Select this option to use the resolution parameters after a lens tune.
•
Tune High Scan Rate
—Select this checkbox to tune the resolution at a scan rate in
addition to the 100 and 1000 amu/s default scan rates.
7. Click the
Targets
tab to configure how you want to tune your targets. Target tuning is
used to adjust the way the ISQ tunes to meet regulatory requirements.