8
Troubleshooting
Investigating Peak Issues
Thermo Scientific
ISQ User Guide
111
Tailing
Sample peaks
Column degradation is
causing activity
Inject a test mixture and evaluate the
column. Replace column if necessary.
Liner is dirty
Clean or replace the liner.
Ion source is dirty
Clean the ion source. Run the method
with higher source temperature, making
certain not to start running samples in
matrix until the source has had several
hours to reach thermal equilibrium.
Glass wool or inlet liner is
causing activity
Replace wool with fresh silanized wool
and install a clean inlet liner.
Inlet temperature is too low Increase the temperature of the inlet.
Column connections are
poor or obstructed
Reconnect the column inlet.
Stationary phase is not
appropriate for your target
compounds
Replace the column and choose a more
appropriate phase for your analysis.
Final hold oven
temperature is too low
Increase the column/oven temperature.
Do not exceed the recommended
maximum temperature for the
stationary phase.
Transferline temperature is
too low
If tailing occurs on late eluting
compounds, it is likely the source or
transferline temperature is too low.
Source temperature is too
low
If tailing occurs on late eluting
compounds, it is likely the source or
transferline temperature is too low.
Poor column
characterization
information about checking for leaks
and column flow.
Table 3.
Troubleshooting Peak Issues in Your Data, continued (Sheet 2 of 3)
Behavior
Characteristic
Cause
Remedy
Summary of Contents for ISQ
Page 1: ...ISQ User Guide 1R120555 0003 Revision D January 2013 ...
Page 22: ......
Page 54: ......
Page 82: ...4 Creating a Method Creating a Method for the GC 60 ISQ User Guide Thermo Scientific ...
Page 124: ......
Page 148: ......
Page 152: ......