2
Changing the Column
Replacing the Factory Installed Column
Thermo Scientific
ISQ User Guide
7
Replacing the Factory Installed Column
To replace the factory-installed column
1. Cool down the GC oven and injector. See the GC documentation for information. After
they are cooled down, power-off the GC.
If you are using hydrogen as a carrier gas, you must cool down and shut off the GC to
prevent the buildup of hydrogen in the vacuum manifold.
2. Click
Shut Down
on the ISQ Dashboard or on the ISQ status page within the
instrument software.
3. Click
Yes
to continue the shutdown process. The high voltages, heaters, and
turbomolecular pump power off. Once the turbomolecular pump reaches 50% speed, or
five minutes elapses, the foreline pump powers off and you may vent the system.
4. Reach around the right side to the back of the instrument and push down on the power
switch to power-off the ISQ instrument.
5. If you are using hydrogen as a carrier gas, unscrew the hydrogen safety screw on the front
door.
Note
The procedure below assumes that optional accessories designed to assist in
changing column without venting the mass spectrometer are not present. For changing
the column when one of these accessories is present, consult the user guide for that
accessory.
Note
If you are running samples, stop the acquisition before powering off the system.
WARNING - BURN HAZARD:
The injector, oven, and transfer line may be hot. Allow
them to cool to room temperature before touching them.
Note
The amber vacuum light on the front of the instrument starts blinking rapidly,
indicating that the mechanical pump has powered off after a five minute period with the
turbomolecular pump off (such as when the instrument is shut down), or due to a
sustained vacuum fault lasting five minutes. When the turbomolecular pump spins down
below 50% speed due to the shut down process, the vacuum light turns off.
WARNING - FIRE HAZARD:
If you are using hydrogen, do NOT reach over the top of the
instrument to power it off. Instead, reach around the right side or go to the back of the
instrument and flip down the power switch.