Weak Sensor Warning
If the sensor that you are calibrating is nearing the end of its useful life, the
Innova alerts you with a Weak Sensor message.
The gas flag will show
“WEAK”
at the conclusion of the calibration period.
R
ESPONDING
TO
A
W
EAK
S
ENSOR
1. If you are manually calibrating the Innova, verify that you are correctly
adjusting the gas concentration to match the concentration of the
calibration gas. If using the Auto Cal method, verify that the gas
concentration in the calibration gas cylinder matches the Auto Cal Gas
Value, see Chapter 5, User Program, Setting Auto Cal Gas Values.
NOTE
A weak sensor is still functioning properly and can continue to be used.
However, it is near the end of its useful life, and should be replaced as
soon as possible, see Chapter 6, Maintenance, Replacing Sensors, for
replacement instructions.