71-6417
21
User Program
E
DIT
A
LARMS
This allows the user to increase or decrease the High and Low Alarms for
each gas. See Chapter 1, Introduction, Table 1-2, Sensor Specifications for
factory default settings.
1. Press
→
to enter the HC screen. The high alarm reading will flash.
2. Press
↑
or
↓
to increase or decrease the alarm setting. If you choose to
keep the alarm as is, press
→
to move to the next alarm setting.
3. Continue step 2 until you have set all gas alarms. After the last gas alarm
is set, the display automatically returns to the “
EDIT ALARMS”
screen.
NOTE
You will scroll through the following gas alarms, appropriate to your
version of Innova: HC, O2, TOXIC1, TOXIC2.
4. Press
↑
or
↓
to continue scrolling through the Advanced Features Menu.
O2 A
LARM
C
ONFIGURATION
This option allows you to set the oxygen alarms (High and Low) to either
increasing or decreasing.
Options you can choose are:
•
ONE INC/ONE DEC
- The warning alarm is set to decreasing at
19.5%VOL and high alarm increasing at 23.5%VOL.
•
TWO ALRM/DECREASE
- Both alarms are set to decrease. Warn at
19.5%VOL and High at 18.0%VOL.
•
TWO ALRM/INCREASE
- Both alarms are set to increase. Warn at
22.5%VOL and high at 23.5%VOL.
The factory default setting is
“ONE INC/ONE DEC”
.
To enter this menu from the Advanced Features Menu:
1. Press
→
to accept
“O2 ALARM CONFIG”
.
2. Press
↓
or
↑
to scroll through the settings. Press the
→
to accept the
setting.
3. Press
↓
to continue scrolling through the Advanced Features Menu.