Alarms
Datalog Alarm
This section describes the alarm indication, display message, and response for
the datalog full alarm. A datalog full alarm occurs when the datalog becomes
full and the datalog full command was accepted as
“STOP ON FULL
”.
Datalog Alarm Indications
Table 3-5 lists the indications for the datalog full alarm.
Responding to the Datalog Alarm
Perform one of the following procedures to reset the datalog alarm.
• Download the data, save and clear the log when prompted.
• Enter the Datalog Features Menu, and update the datalog full
command to
“OVRWRITE ON FULL
”, see Chapter 4, User
Program, Log Memory Manager option.
• Enter the Datalog Features Menu and clear the datalog, see Chapter 4,
User Program, Clear Log Memory.
Temperature Alarm
The Innova has an alarm to alert the user that the unit is being operated at a
temperature that exceeds its operating specification of -20°C to +45°C (-4°F
to +113°F). If the temperature is outside this range, the following message
appears every 30 seconds and the sensor readings may be inaccurate:
“TEMP
WARNING”
alternating with the gas readings.
The buzzer beeps and lights flash twice every 30 seconds, and the condition
will clear automatically when the temperature returns to within the operating
limits.
NOTE
The temperature alarm is user enabled or disabled, with the default
setting “disabled”, see Chapter 4, User Program, Temperature Alarm.
.
Table 3-5
Alarm Indications (Datalog Full Alarm)
Alarm
Display
Light/Buzzer
Datalog Full
LOGGING STOPPED flashing/pulsing (every 30 seconds)