28
71-6417
Innova Series Operator’s Manual
Datalog Features Menu
The Datalog Features Menu allows you to change the following parameters of
the instrument.
The menu is as follows:
• Adjust Date and Time
• Edit Operator IDs
• Edit Location IDs
• Edit Instrument ID
• Log Memory Manager Option
• Log Averages Selection
• Log Interval Selection
• View Log Status
• Clear Log Memory
• View Log Began
• Exit
To enter the Datalog Features Menu:
1. Press and hold both the
↑
AIR
and
↓
RANGE
buttons for 3 seconds.
The screen displays
“FEATURES MENU”
and counts down from 3
seconds.
2. The first screen displayed after the countdown is
“BEGIN CAL”
.
Press the
↓
button twice. The display will read
“DATALOG
MENUS”
.
3. Press the
→
button. If prompted by
“LOGWORD”
and
“***”
, enter
a LogWord as described earlier in this chapter. When entered correctly,
or if LogWord is not enabled, you are now at the first option
“ADJUST DATE AND TIME”
.
NOTE
The Innova has a time-out feature. While in the Advanced Features
Menu, if a button is not pressed within 10 seconds the display will flash
and the buzzer will beep for 10 seconds. If no button is pressed during
the 10 seconds, the Innova will return to normal operating mode.