71-6417
27
User Program
E
NTER
/E
DIT
C
AL
W
ORD
, P
ASS
W
ORD
,
OR
L
OG
W
ORD
These options add extra security to the calibration process (CALWORD),
Advanced Features Menu (PASSWORD), and Datalog Features Menu
(LOGWORD). By changing the default factory setting of
“***”
, you
automatically turn this option on, and password protect that area of the user
program. If turned on, a code must be entered before one may continue a
process. If “***” is entered, the password security is turned off. The factory
default setting for each password is “***”, or disabled.
To add or change the code from the Advanced Features Menu,
1. Press
→
at the appropriate screen
“CHANGE CALWORD”
,
“CHANGE PASSWORD”
or
“CHANGE LOGWORD”
. The
screen will display either three asterisks “***”, or the three digit code,
with the cursor on the first character.
2. Press
↓
or
↑
to change the first character of the code. You have the
option of
“*, 0 through 9”
.
NOTE
When entering a CalWord, PassWord, or LogWord all three characters
must be a number, 0 through 9. A password containing one or more “*”
is not valid, no security protection is implemented if a password
contains an “*”.
3. Press
→
to accept the setting, and move on to the second character. If
you do not want to change the first character, press
→
to move on to the
second character.
4. Repeat the process through the third character.
5. Press
↓
to continue scrolling through the Advanced Features Menu.
E
XIT
The
‘EXIT’
screen exits you from the Advanced Features Menu. Press
→
at
“EXIT”