32
71-6417
Innova Series Operator’s Manual
L
OG
S
TATUS
This is a display screen only, which indicates the number of datalog hours or
days left in memory that is not currently occupied.
To display the number of hours or days left, from the Datalog Features Menu:
1. Press
→
at the screen
“LOG STATUS”
. The screen displays reads
“XX LOG HRS LEFT”
or
“XX LOG DAYS LEFT”
(“XX” for
number of hours or days available in memory). The display will
automatically go back to the
“LOG STATUS”
screen.
2. Press
↑
or
↓
to continue scrolling the Datalog Features Menu.
C
LEAR
L
OG
To clear the data in the memory, from the Datalog Features Menu:
1. Press
→
at the screen
“CLEAR LOG”
. The screen displays
“
↑
AND
↓
TO CLEAR”
.
2. Press and hold
↑
and
↓
for 3 seconds and release. The display reads
“CLEARING LOG”
and beeps for several seconds. When complete
the display indicates
“LOG CLEARED”
.
NOTE
If you do not want to clear the log, press the
→
button and continue
scrolling through the Datalog Features Menu.
CAUTION
You cannot recover data after you clear the log. Download all
required data before you clear the log.
L
OG
B
EGAN
This screen displays the date and time when the log was last cleared.
To display this data and time, from the Datalog Features Menu:
1. Press
→
at the screen
“LOG BEGAN”
. The screen displays the data
and time. For example, “7/12/99 10:00:00”. The display automatically
goes back to the
“LOG BEGAN”
screen.
2. Press
↑
or
↓
to continue scrolling on the Datalog Features Menu.
E
XIT
The ‘EXIT’ screen exits you from the Datalog Features Menu. Press
→
to exit
the menu and return to the normal operating mode.