1
Getting Familiar with your EA IsoLink IRMS System for CNSOH
Connections Panel
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
21
Connections Panel
The connections panel is shown in
Figure 19.
View of the Connections Panel
The connections panel is subdivided into the following connecting areas:
interface
,
power supply
, and
gases supply
.
Interface Area
The
interface
area comprises the connectors for connecting the autosamplers and the helium
management device, and the automatic switching valve. See
.
Figure 20.
Interface Area
• A 9-pin connector marked
RS 232
to dialog with the computer via serial line.
• A 25-pin connector marked
Aux Connector
for the remote start of the instrument.
• A 2-pin connector marked
Autosampler HO
for the MAS Plus autosampler installed on the
left
channel
for HO determinations.
• A 2-pin connector marked
Autosampler NCS
for the MAS Plus autosampler installed on the
Right channel
for NCS determinations.
• A 9-pin connector marked
Trigger Input
for the control of the automatic switching valve.
Power Supply
Gas Inlets