Maintenance
Current Maintenance Program
126
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Figure 137.
Graphite Crucible Removal
1. Cool the furnace to a temperature below
600 °C
.
2. Put the ConFlo in standby or close needle valve to IRMS.
3. After reaching the temperature switch
Off
the flow of helium. Wait approximately five minutes to
depressurize the system.
4. Remove the autosampler.
5. Carefully remove the graphite tube (for example with a pair of tweezers), see
A
, and
place it on a clean cloth.
6. Insert the crucible remover inside the graphite crucible. See
B
.
7. Pull the graphite crucible out of the reactor. See
C
in
8. Drop a new graphite crucible into the reactor, or remove the deposits outside and inside of the
graphite crucible and use it again.
9. Remount the parts proceeding in the reverse order.
Maintenance measures for pyrolysis unit
To facilitate proper work, adhere to the described maintenance recommendations and procedures.
This will also
increases the life time of the pyrolysis reactor and the furnace heater.
A
Graphite Tube
Graphite Crucible
B
C
Crucible Remover Tool