Maintenance
Maintaining the MAS Plus Autosampler
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
129
Maintaining the MAS Plus Autosampler
The MAS Plus autosampler does not normally require maintenance. However, when the instrument is
extensively used, it is a good practice to clean from time to time the shaft housed in the autosampler
operating as described in the following operating sequence.
To clean the shaft of the MAS Plus autosampler
1. In Isodat Software Suite, right click in the Flash IRMS dialogue box and then select settings for
EA.
2. In Eager
Smart
Data Handling Software Main Menu
Edit | Edit Elemental Analyzer parameters
,
or just click the
icon. The
Analyzer Parameter
window is visualized.
3. Click
Send
and wait 2-3 minutes to discharge the gas out the circuit.
4. Remove the frontal cover of the MAS Plusautosampler as shown in
Figure 140.
Removal of the MAS Plus Autosampler Cover
5. From Main Menu select the menu
View | View Elemental Analyzer status
,
or
click the icon
.
6. In the
Status
window select the
Special function
tag,
then click
Step sampler tray position
.
The autosampler mechanism pushes the shaft forward and then ejects it.
7. Take out the shaft from the autosampler, as shown in
.
Materials Required
Phillips screwdriver
Cloth
Silicon grease (For use at pressures down to 10
-6
mm Hg)
Cover