Installing AI 1310/AS 1310 Autosampler
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
79
4. Check the correct alignment of the sampling system support, then fix it by tightening the proper
fixing screws.
5. Insert the sample tray into the sampling unit base.
•
AI 1310
— Insert the 8-position sample tray into the appropriate housing of the sampling
unit base. See
Figure 80.
8-position Sample Tray
•
AS 1310
— Insert the dedicated support plate into its appropriate housing of the sampling
unit base. Place the 105-position sample tray on the hub located on the support. The system
will automatically recognize the sample tray at the instrument power on. See
.
Figure 81.
105-position Sample Tray
Installing the Syringe
The installation of the syringe is a simple operation. However, it must be performed with caution to
avoid damages to the syringe needle and ensure an optimal performance of the injection device.
The standard syringes have 10
μ
L and 250
μ
L capacity with a 50 mm needle. It is also possible to
install 50
μ
L and 100
μ
L syringes with needles of 50 mm.