Getting Familiar with your EA IsoLink IRMS System for CNSOH
Electrical Compartment
20
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Main Voltage Compartment
It contains the mains power circuits and the Safety Cut Off device.
details the function of each component present in the mains voltage section.
TCR 1112
Operates and controls the furnaces thermoregulation.
PWR 1112
Receives voltage supplies from the TRF 1112 transformers board.
Generates voltage supply for the electronic control circuits.
FP 1112
Synoptic panel
Table 4.
Description of the Function of the Electronic Boards (Sheet 2 of 2)
Board
Function
Table 5.
Description of the components of the Main voltage compartment
Component
Description
TRF 1112 Transformers
Board
Receives the mains power and supplies it to the following devices:
• Cooling fan
• Furnaces transformers
• Heater of the detector thermostatic chamber
Six fuses are provided on the board. See
AC 1112 Furnaces
Power Supply
Supplies 48 Vac power to the furnaces. It contains the SSR relays for the
furnaces control. Also see the section
“Devices for the Furnaces Control”
. Two fuses are provided on the board. See
Breaker
Instrument On/Off main switch.
Table 6.
Fuses of the High voltage compartment
Board
Fuse
Type
Protection
TRF 1112
F1
F1A; IEC 127/I (5 x 20 mm)
Power supply to LEFT and RIGHT
furnaces transformers
F2
F0.315A; IEC 127/I (5 x 20 mm)
Fan
F3
F1.6A; IEC 127/I (5 x 20 mm)
Main power (Breaker)
F4
F1A; IEC 127/I (5 x 20 mm)
LEFT and RIGHT furnaces
transformers
F5
F0.315A; IEC 127/I (5 x 20 mm)
Fan
F6
F1,6A; IEC 127/I (5 x 20 mm)
Mains power (Breaker)
AC 1112
F1
FF12 A; IEC 269 (1.3 x 38 mm)
LEFT Furnace power circuit
F2
FF12 A; IEC 269(1.3 x 38 mm)
RIGHT Furnace power circuit