Devices for the Furnaces Control
describes the function of each device:
Devices Supplying the Furnaces
details the function of each device:
Status Panel
The LED status panel shows the instrument operating conditions, and it is located on the right side of
the instrument front panel. See
Figure 25.
LED Status Panel
Each LED lights up when the relevant function is active.
illustrates the meaning of each
function:
Table 9.
Description of the Devices Controlling the Furnaces
Device
Function
LTA-1 LEFT
LTA-1 RIGHT
They read the values of the thermocouple present in the relevant furnace and
send the signals to the
TCR 1112
board.
SSR LEFT
SSR RIGHT
Solid State Relays (SSR) contained in the
AC 1112
board.
Each SSR is coupled with a proper safety sensor, which detects any malfunction.
The SSR control the power supply to the relevant furnace and cut off power to
the heating resistor when the thermocouple detects temperature values
exceeding the setpoint.
Table 10.
Description of the Devices Supplying the Furnaces
Device
Function
T1 Transformer
Supplies 48 V voltage to the right furnace resistor. It is provided with a safety
thermal protection, which cuts off power in case of overheating.
T2 Transformer
Supplies 48 V voltage to the left furnace resistor. It is provided with a safety
thermal protection, which cuts off power in case of overheating.