6
Using EA IsoLink IRMS System for CNSOH
Setting Up a HO Method
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
99
The
MethodSwicher
window opens, see
.
Figure 107.
Switching Between Methods
9. For
Gasconfiguration
, select the gas configuration to be switched to, for example,
CO
.
10. Select the
Event
when the switch is to be performed, for example,
Isodat@Eval@PeakFound
.The
switch is performed when the peak is found.
11. For
Waiting Time
, enter for example, after 30 s.
If a peak is found, the jump takes place. If no peak is found, the waiting time elapses before the
jump takes place immediately.
12. Click
OK
.
13. Click the tab
Evaluation@H2
.
Figure 108.
Evaluation@H2 Page of the Dual Measurement Method
Note
There are Evaluation pages, Peak Detection pages and Printout pages are for each, H
2
and CO, for example, Evaluation@H2 and Evaluation@CO2.