7
Maintenance
Upgrading EA for 25 mm OD Macro Reactor
134
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
5. On the top panel of the instrument remove the left/right upper connector for the 18 mm OD
reactor unscrewing the two fixing Allen screws with the proper Allen wrench. Place and fix the
upper connector for 25 mm OD macro reactor.
6. On the top of left/right furnace remove the adaptor for 18 mm OD reactor and its sealing adapter.
7. Reinstall the protecting plate and the cover of the furnace compartment.
8. Install the macro reactor tube into the furnace and reinstall the MAS Plus.
Upper Connector for
18 mm OD Reactor
Upper Connector for
25 mm OD Macro Reactor
Adaptor for 18 mm
OD Reactor
Sealing Adapter