Installation
Installing the Sampling System Support on the TRACE 1300/TRACE 1310
Thermo Scientific
AI 3000/AS 3000 Sampling System - Operating Manual
41
Figure 36.
Assembling (2)
3. Guide the three fixing screws into the corresponding fixing holes located on the GC top.
See A’, B’, and C’.
4. Tighten the three fixing screws.
5. If not already done, mount and fix the guide pivot on each top plate by using the proper
fixing screw.
6. Place the two top plates over the double bottom plate.
7. Guide each retainer and its fixing screw into the sliding slot present on each top plate.
8. Align each retainer fixing screw into the corresponding fixing hole located on the double
bottom plate.
9. Tighten each retainer fixing screw to block the corresponding upper plate to the double
bottom plate.