3
Installation
Sampling System Support
Thermo Scientific
AI 3000/AS 3000 Sampling System - Operating Manual
23
Sampling System Support
The AI 3000/AS 3000 II is installed on the GC using the appropriate support provided.
According to your GC, please refer to:
•
Sampling System Support for TRACE 1300/TRACE 1310
•
Sampling System Support for TRACE GC Ultra
•
Sampling System Support for FOCUS GC
Sampling System Support for TRACE 1300/TRACE 1310
The AI 3000/AS 3000 II can be installed only on the
right
side of the GC on the
front
or
back
injector module or on both when in
Gemini
configuration by using the proper sampling
system support.
The sampling unit support consists of a bottom and a top plates.
1. The bottom plate can be
L-shape
(see both front/back installation) or
double
(see
Gemini installation). The plate is provided with a series of holes for the corresponding
fixing screws and must be fixed on the GC.
2. The movable top plate is mounted over the bottom plate and it is provided with a guide
pivot for the accommodation and centering of the sampling unit. The movable top plate
can slides over the bottom plate through a sliding slot and a retainer.
This allows moving away the injection device of the sampling unit from the injector
module when the latter will require maintenance.
To use the proper sampling system support on the TRACE 1300/TRACE 1310 please refer
to:
•
Support for the Installation on the Front/Back Injector Module
•
Support for the Installation in Gemini Configuration
•
Support for the Installation When a TSQ Quantum Mass Spectrometer is Coupled with
the GC
Note
The FOCUS GC MS Version, may accept the AI 3000/AS 3000 II also on the left
side. Please refer to
“Assembling of the Sampling System Support on FOCUS GC MS
To reduce the FOCUS GC working area, the AS 3000 may be installed as described in
.
Note
After the installation of the sampling unit on the proper sampling system
support, it is possible locking its plates if desired.