4
AI 3000/AS 3000 II Control
Control Through the TRACE GC Ultra
Thermo Scientific
AI 3000/AS 3000 Sampling System - Operating Manual
63
AI 3000/AS 3000 II Menu
To set the parameters of the AI 3000/AS 3000 II method, press
Autosampler
key to open the
relevant
menu described in
Table 8.
Control Through the TRACE GC Ultra: Autosampler Menu
Menu
Range
Comment
AUTOSAMPLER
This line displays the menu title.
Sample volume
0–5 μL
with increments of
0.1 μL
This parameter allows to set the sample
quantity to be injected.
Sample rinses
0–15
This parameter allows to set the number of
syringe pre-washings with the sample.
Plunger strokes
0–15
This parameter allows to set the number of
plunger strokes to eliminate air bubbles
forming during the sample withdrawal.
Pre wash solvent
A, B, C, D
A+B
C+D
This parameter allows to select the vial, or
combination of two vials, containing the
washing solvent.
Press
Mode/Type
to select the solvent vial
to be used.
Pre wash cycles
0–15
This parameter allows to set the number of
syringe pre-washings with the selected
solvent.
Post wash solvent
A, B, C, D
A+B
C+D
This parameter allows to select the vial, or
combination of two vials, with the washing
solvent.
Press
Mode/Type
to select the vial with the
solvent to be used.
Post wash cycles
0–15
This parameter specifies the number of
syringe post-washings with the solvent
selected.
Extended control
See
This parameter allows to set optional
injection parameters.
Press
Mode/Type
t
o
enter the submenu.
When no vial abort
See
This line shows the menu title.