4
AI 3000/AS 3000 II Control
Control from PC
58
AI 3000/AS 3000 Sampling System - Operating Manual
Thermo Scientific
Washes
This field contains parameters controlling syringe washing with solvents and the sample.
Pre-Injection
These parameters are applied to pre-injection washing with a solvent chosen from the vials in
the Washing and Waste tray. See
for details.
Sample
This is the number of times that the syringe is rinsed with the sample prior to injection. Select
a value in the range 1-15.
Post-Injection
These parameters are applied to post-injection washing with a solvent chosen from the vials in
the Washing and Waste tray. See
for details.
Post-injection Dwell
Time
This is the time that the needle is held in the inlet after the
injection. Enter a value in the range 0-63 s.
Table 4.
Injection Parameters (Sheet 2 of 2)
Parameter
Description
Table 5.
Washer Parameters
Parameter
Description
Pre-Injection
These parameters are applied to pre-injection washing with a
solvent chosen from the vials in the Washing and Waste tray. See
for details.
Sample
This is the number of times that the syringe is rinsed with the
sample prior to injection. Select a value in the range 1-15.
Table 6.
Pre-injection parameters
Item
Description
Solvent
Choose a solvent from the drop-down list. The options are:
A
,
B
,
C
,
D
,
A+B
and
C+D
. These settings correspond to the vials in the
Washing and Waste tray.
Cycles
Select the number of times the solvent is cycled through the
syringe. The range is 1-15.