AI 3000/AS 3000 II Control
Control Through the FOCUS GC
66
AI 3000/AS 3000 Sampling System - Operating Manual
Thermo Scientific
Control Through the FOCUS GC
This paragraph describes how to program and control the AI 3000/AS 3000 II through the
FOCUS GC keypad.
The AS 3000 can automatically analyze groups of samples under different analytical
conditions when a
sequence
of samples is programmed.
The sequence includes a set of information for the group of samples and their position on the
sample tray. The sequence also specifies the method to be used to process each group of
samples. Refer to paragraph
Sequence of Samples with the AS 3000
Configuring the AI 3000/AS 3000 II
When the GC is switched on, the presence and type of sampling system installed is
automatically acknowledged.
To configure the AI 3000/AS 3000 II:
1. In Main Menu move the cursor until
Others
is displayed, then press
Enter
.
2. Scroll to select
Configuration
, then press
Enter
.
3. Scroll to
Autosampler
, then press
Enter
.
The following message will be displayed according to the AI 3000/AS 3000 II installed:
for the AS 3000.
or
for the AI 3000.
The parameter
Syringe volume
can be modified according to the syringe volume, whereas the
parameter
Sample
tray is not editable.
CONFIG AUTOSAMPLER
Syringe Volume
Sample tray
10
105
CONFIG AUTOSAMPLER
Syringe Volume
Sample tray
10
8