1
Introduction
The AI 3000/AS 3000 II Sampling System
Thermo Scientific
AI 3000/AS 3000 Sampling System - Operating Manual
3
Figure 2.
AS 3000 Auto-sampler
Sampling Unit
The sampling unit consists of the following major components:
Base
This is the structure bearing the sampling unit components.
Turret
It consists of a vertical moving structure that accommodates the injection unit and
accomplishes the necessary movements to carry out the three main operating steps: sample
drawing, injection, syringe washing. The turret is provided with a safety door allowing access
to the syringe.
Turret
Safety Door
Injection Unit
Centering Plate
Washing and Waste Tray
Base
105-position Sample Tray