4
AI 3000/AS 3000 II Control
Control from PC
Thermo Scientific
AI 3000/AS 3000 Sampling System - Operating Manual
57
This page includes the following fields, menus and messages:
•
•
•
•
•
•
•
Sampling
Set up the sampling parameters in this field. See
. for details.
Injection
These parameters control the behavior of the syringe when the needle is inserted into the
inlet. See
for details.
Table 3.
Sampling Parameters
Parameter
Description
Sample Volume
Enter a value for the injection volume. The range depends on the
size of syringe selected in the Instrument Configuration. For a 5 μl
syringe the range is 0.0-2.5 μl and for a 10 μl syringe, 0.0-5.0 μl.
Plunger Strokes
Select the number of strokes of the plunger used to eliminate air
bubbles. The range is 1-15.
Viscous Sample
Select the check box to indicate that the sample is viscous. Clearing
the check box implements a fast aspiration speed while a selected
check box leads to a slow one.
Sampling Depth
Choose the position of the needle in the vial from which the
aliquot is drawn. The options are
Bottom
or
Center
.
Table 4.
Injection Parameters (Sheet 1 of 2)
Parameter
Description
Injection Depth
Choose an injection depth. The options are
Standard
or
Minimum
. If the pre and post -injection dwell times are set to
zero, the autosampler defaults to the
Minimum
setting.
Pre-injection Dwell
Time
This is the time that the needle is held in the inlet prior to
injection. Enter a value in the range 0-63 s.