Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 43i Trace Level-Enhanced Instruction Manual
3-43
In the Main Menu, choose Instrument Controls > IO Configuration >
Analog Output Config > Select Channel >
Set Minimum
or
Maximum
Value
.
MINIMUM OUTPUT PERCENT:
SELECTED OUTPUT: V 1
CURRENTLY: N/A %
SET TO: 0000.
5
% ?
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Table 3–6.
Analog Output Zero to Full Scale Values
Output
Zero % Value
Full-Scale 100% Value
SO2
Zero (0)
Range Setting
LO SO2
Zero (0)
Range Setting
HI SO2
Zero (0)
Range Setting
Range SO2
Recommend not to change the setting for this output
Internal Temp
User-set alarm min value
User-set alarm max value
Chamber Temp
User-set alarm min value
User-set alarm max value
Perm Oven Gas Temp
Perm Oven Gas alarm min value
Perm Oven Gas alarm max value
Perm Oven Heater
Temp
Perm Oven Heater alarm min
value
Perm Oven Heater alarm max
value
Pressure
User-set alarm min value
User-set alarm max value
Sample Flow
User-set alarm min value
User-set alarm max value
PMT Volts
700 volts
1100 volts
Flash Voltage
User-set alarm min value
User-set alarm max value
Flash Reference
User-set alarm min value
User-set alarm max value
Everything Else
0 units
10 units
The Choose Signal To Output screen displays a submenu list of the analog
output signal group choices. Group choices are Concentrations, Other
Measurements, and Analog Inputs (if the I/O expansion board is installed).
This allows the user to select the output signal to the selected output
channel. In dual or auto range mode, “HI” or “LO” is displayed to indicate
high or low concentrations. Range (SO
2
) is visible only in auto range mode.
The Concentrations screen is shown below. See Table 3-8 that follows for a
list of items for each signal group choice.
Choose Signal To Output