C-Link Protocol Commands
Measurement Configuration
B-28
Model 43
i
Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
Send:
range
Receive:
range 6: 5000E-2 ppm
set range
Selection
set high range
Selection
set low range
Selection
These commands select the SO
2
full-scale ranges, according to
Table B–6
and
Table B–7
. The example that follows sets the SO
2
full-scale range to
100 ppm.
Send:
set
range
3
Receive:
set range 3 ok
Table B–6.
Standard Ranges
Selection
ppm
mg/m
3
0 0.05
0.2
1 0.1
0.5
2 0.2
1
3 0.5
2
4 1
5
5 2
10
6 5
20
7 10
25
8 C1
C1
9 C2
C2
10 C3
C3
Table B–7.
Extended Ranges
Selection
ppm
mg/m
3
0 0.5
2
1 1
5
2 2
10
3 5
20
4 10
50
5 20
100
6 50
200