Operation
Service Menu
Thermo Fisher Scientific
Model 43i Trace Level-Enhanced Instruction Manual
3-77
PERM OVEN SETPOINT:
CURRENTLY: NOT PRESENT
SET TO: 45
o
C ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Factory Calibrate Gas Thermistor submenu is used for calibrating the
permeation oven gas thermistor to either low point, high point, or set
defaults.
In the Main Menu, choose Service > Perm Oven Settings >
Factory Cal
Gas Therm
.
FACTORY CAL OVEN THERM:
>LOW POINT
HIGH POINT
SET DEFAULTS
RANGE AVE DIAGS ALARM
The Calibrate Gas Thermistor Low Point screen allows the user to view
and set the permeation oven thermistor resistance. The low and high point
screens function the same way.
In the Main Menu, choose Service > Perm Oven Settings > Factory Cal
Gas Therm >
Low Point
.
CAL GAS THERM LOW POINT:
CURRENTLY: 3850 Ohms
SET TO: 0400
0
Ohms
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Set Defaults screen allows the user to reset the configuration values to
factory defaults.
In the Main Menu, choose Service > Perm Oven Settings > Factory Cal
Gas Therm >
Set Defaults
.
Factory Calibrate Gas
Thermistor
Low and High Points
Set defaults