![Thermo Scientific 43i Instruction Manual Download Page 65](http://html.mh-extra.com/html/thermo-scientific/43i/43i_instruction-manual_1105147065.webp)
Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 43i Trace Level-Enhanced Instruction Manual
3-27
CONCENTRATIONS:
>NONE
SO2 <--
LO SO2
HI SO2
RANGE (SO2)
RANGE AVG DIAGS ALARM
The Other Measurements screen allows the user to select the output signal
that is tied to the selected field item. The selected item is shown by “<--”
after it. Items displayed are determined by the options installed. Note that
at this point, pressing
indicates that these are proposed changes as
opposed to implemented changes. To change the selected record format
and erase record log file data, see “Commit Content” that follows.
In the Main Menu, choose Instrument Controls > Datalogging Settings >
Select Content > Select Field >
Other Measurements
.
OTHER MEASUREMENTS:
>NONE
INT TEMP <--
CHAMBER TEMP
PERM OVN GAS
PERM OVN HTR
CHAMBER PRES
SAMPLE FLOW
ª
RANGE AVG DIAGS ALARM
The Analog Inputs screen allows the user to select the output signal (none
or analog inputs 1-8) that is tied to the selected field item. The selected
item is shown by “<--” after it. Note that at this point, pressing
indicates that these are proposed changes as opposed to implemented
changes. To change the selected record format and erase record log file
data, see “Commit Content” that follows.
In the Main Menu, choose Instrument Controls > Datalogging Settings >
Select Content > Select Field >
Analog Inputs
.
Other Measurements
Analog Inputs