Operation
Service Menu
3-70
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
In the Main Menu, choose Service > Flow Calibration >
Zero
.
CALIBRATE FLOW ZERO:
CURRENTLY: 0.000 LPM
SET TO: 0.000 LPM ?
DISCONNECT PUMP AND
SAVE ZERO PRESSURE
RANGE AVG DIAGS ALARM
The Calibrate Flow Span screen allows the user to view and set the flow
sensor calibrate span point.
Note
An independent flow sensor is required to read the flow, then the
operator enters the flow value on this screen to perform the calibration.
▲
In the Main Menu, choose Service > Flow Calibration >
Span
.
CALIBRATE FLOW SPAN:
CURRENTLY: 0.508 LPM
SET TO: 0.
8
08 LPM ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Restore Default Flow Calibration screen allows the user to reset the
configuration values to factory defaults.
In the Main Menu, choose Service > Flow Calibration >
Set Defaults
.
Calibrate Flow Span
Restore Default Flow
Calibration
RESTORE DEFAULT CAL:
RESTORE
RANGE AVG DIAGS ALARM
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE