Thermo Fisher Scientific
Model 43i Trace Level-Enhanced Instruction Manual
2-1
Chapter 2
Installation
Installation of the Model 43
i
Trace Level-Enhanced includes lifting the
instrument, unpacking and inspection, connecting sample, zero, span, and
exhaust lines, and attaching the analog outputs to a recording device. The
installation should always be followed by instrument calibration as
described in the “Calibration” chapter of this manual.
This chapter provides the following recommendations and procedures for
installing the instrument:
●
Lifting on page 2-1
●
Unpacking and Inspection on page 2-1
●
Setup Procedure on page 2-3
●
Connecting External Devices on page 2-5
●
Startup on page 2-9
When lifting the instrument, use a procedure appropriate to lifting a heavy
object, such as, bending at the knees while keeping your back straight and
upright. Grasp the instrument at the bottom in the front and at the rear of
the unit. Although one person can lift the unit, it is desirable to have two
persons lifting, one by grasping the bottom in the front and the other by
grasping the bottom in the rear.
Equipment Damage
Do not attempt to lift the instrument by the cover or
other external fittings.
▲
The Model 43
i
Trace Level-Enhanced is shipped complete in one
container. If there is obvious damage to the shipping container when you
receive the instrument, notify the carrier immediately and hold for
inspection. The carrier is responsible for any damage incurred during
shipment.
Use the following procedure to unpack and inspect the instrument.
Lifting
Unpacking and
Inspection