Operation
Service Menu
3-78
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
RESTORE DEFAULT SETTINGS:
FOR PERM OVEN GAS THERM
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
The Factory Calibrate Oven Thermistor submenu is used for calibrating
the permeation oven heater thermistor to either low point, high point, or
set defaults.
In the Main Menu, choose Service > Perm Oven Settings >
Factory Cal
Oven Therm
.
FACTORY CAL OVEN THERM:
>LOW POINT
HIGH POINT
SET DEFAULTS
RANGE AVE DIAGS ALARM
The Calibrate Oven Thermistor Low Point screen allows the user to view
and set the permeation oven thermistor resistance. The low and high point
screens function the same way.
In the Main Menu, choose Service > Perm Oven Settings > Factory Cal
Oven Therm >
Low Point
.
CAL OVEN THERM LOW POINT:
CURRENTLY: 3850 Ohms
SET TO: 0400
0
Ohms
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Factory Calibrate Oven
Thermistor
Low and High Points