Operation
Service Menu
3-76
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
CAL GAS THERM (BATH):
CURRENTLY: 45.80
o
C
SET TO: 45.0
0
o
C
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Calibrate Gas Thermistor Resistance screen allows the user to view and
set the permeation oven gas thermistor resistance to a known resistor value.
CAL GAS THERM (RESISTOR):
CURRENTLY: 3850 Ohms
SET TO : 0400
0
Ohms
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Calibrate Oven Thermistor menu is used to view and set the
permeation oven thermistor resistance to a known resistor value.
In the Main Menu, choose Service > Perm Oven Settings >
Cal Oven
Thermistor
.
CAL OVEN THERM
(RESISTOR):
CURRENTLY: 3850 Ohms
SET TO : 0400
0
Ohms
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Permeation Oven Setpoint screen is used to select the permeation oven
state to not present, or the desired temperature choice of 30, 35, and 45
°
C.
In the Main Menu, choose Service > Perm Oven Settings >
Perm Oven
Selection
.
Resistor
Calibrate Oven
Thermistor
Permeation Oven
Setpoint