C-Link Protocol Commands
Measurement Configuration
Thermo Fisher Scientific
Model 43
i
Trace Level-Enhanced Instruction Manual
B-29
Selection
ppm
mg/m
3
7 100
250
8 C1
C1
9 C2
C2
10 C3
C3
custom
range
range
= | 1 | 2 | 3 |
This command reports the user-defined value of custom
range
1, 2, or 3.
The example that follows reports that custom range 1 is defined to 5.50
ppm.
Send:
custom
1
Receive:
custom 1 550E-2 ppm
set custom
range
range
value
set custom 1 range
value
set custom
2 range
value
set custom 3 range
value
These commands are used to set the maximum concentration for any of the
three custom
ranges
1, 2, or 3 to range
value
, where
value
is a floating-point
number representing concentration in ppm or mg/m
3
. The example that
follows sets the custom 1 range to 55.5 ppm.
Send:
set
custom
1
range
55500
Receive:
set custom 1 range 55500 ok
range mode
This command reports the current range mode.
Send:
range
mode
Receive:
range mode single
set range mode
mode
This command sets the current range mode to single, dual, or auto. The
example that follows sets the range mode to single.
Send:
set
range
mode
single
Receive:
set range mode single ok
gas mode
This command reports the current mode of sample, zero, or span. The
example that follows reports that the gas mode is sample.