Operation
Service Menu
3-74
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
The Analog Input Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog input.
In the Main Menu, choose Service > Analog Input Calibration > Select
Channel >
Calibrate Zero
. (Hook up a voltage source of 0 V to the analog
input channel.)
ANALOG INPUT CAL: ZERO
DISCONNECT SELECTED INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 0.00 V ?
CALIBRATE INPUT TO ZERO
RANGE AVG DIAGS ALARM
The Analog Input Calibration Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog input.
In the Main Menu, choose Service > Analog Input Calibration > Select
Channel >
Calibrate Full Scale
. (Hook up a voltage source of 10 V to the
analog input channel.)
ANALOG INPUT CAL: SPAN
PROVIDE VOLTAGE TO INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 9.80 V
SET TO: 10.00 V ?
CALIBRATE TO VALUE
RANGE AVG DIAGS ALARM
Analog Input Calibration
Zero
Analog Input Calibrate
Full-Scale
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5
INPUT CHANNEL 6
INPUT CHANNEL 7
ª
RANGE AVG DIAGS ALARM
INPUT CHANNEL 8
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM