set date
mm-dd-yy
mm =
month
dd =
day
yy
=
year
This command sets the date of the analyzer’s internal clock. The example
that follows sets the date to December 1, 2004.
Send:
set
date
12-01-04
Receive:
set date 12-01-04 ok
set default params
This command sets all the parameters to their default values. This does not
affect the factory-calibrated parameters.
Send:
set
default
params
Receive:
set default params ok
lamp status
This command reports the status of the flash lamp on or off. The example
that follows reports the flash lamp is on.
Send:
lamp
status
Receive: lamp
status
on
set lamp
onoff
These commands set the flash lamp
on
or
off
. The example that follows
turns the flash lamp off.
Send:
set
lamp
off
Receive:
set lamp off ok
pmt status
This command reports the status of the PMT on or off. The example that
follows reports that the PMT is on.
Send:
pmt
status
Receive:
pmt status on
set pmt
onoff
These commands set the PMT
on
or
off
. The example that follows turns