95
Care has to be taken at first installation, as WIN 3000 needs proper installation
on the drive PC. The setting of the interfaces needs to be done properly. For
this consult the documents in PDF format available on the CD delivered with
the instrument.
601-326B - NSG unit & WIN 3000 Installation - Quick installation guide english.pdf
Refer to sections 1, 2, 3, 4 and 5 of this document.
The factory setting of NSG 3040_ERC series is IP address 10.10.10.10, SubNet
255.0.0.0, Port 1025
WARNING - The factory setting of NSG 3040_ERC series
is IP address 10.10.10.10, SubNet 255.0.0.0, Port 1025.
Care has to be taken to remember the new settings if
these get changed.
Forgotten IP settings can only be resetted in Teseq
service centers. Therefor it is strongly recommended
to leave the factory set IP and to install INA 3011 option
on the user PC. The second IP address dedicated to
the NSG 3040 control may be set to the fix IP address
10.10.10.11 (refer to quick installation guide) and so will
not interfere with the PC network settings.