135
Once the right Volts to A/m ratio – which is available on the INA 752 as well then
in its calibration certificate is entered in its respective field, the user will be able
to enter the required test level directly in A/m – no need for him to proceed
with calculations, the instrument software does the job.
In case an INA 702 loop antenna is used, the termination
plug labelled “Pulse“ needs to be used.
Technical specifications NSG 3040 (CWM 3450) - INA 752 - INA 701/
INA 702
Parameter
Value
Volts to A/m ratio*:
3.55
Magnetic field adjustment:
Software driven, NSG 3040 settings
200 to 4400 V => allows 60 to 1200 A/m
INA 752 weight:
0.6 kg
INA 752 dimensions:
140 x 75 x 55 mm
* Typical, exact value is given on INA 752 front panel
Following test screen will appear: