119
1) Verify the setting of input voltage selector and adjust it to the right mains
voltage value if required
2) Connect instrument power from the mains
3) Remove 25 way Sub D plug at rear of NSG 3040
4) Connect this connector to X2 of INA 6502
5) Connect master controller 25 way output to INA 6502 X1 plug, using
system interface cable delivered with INA 6502
6) Connect INA 6502 - EUT power out to NSG 3040 – EUT power input
7) Connect INA 6502 - EUT power in to mains using EUT power in cable
delivered with NSG 3040
The equipment should be switched off during installa-
tion and interconnection.
Because of the capacitors in the internal coupler of NSG
3040, earth leakage currents of up to 4 A can occur in
the EUT power supply network. The test system must
therefore be correctly earthed and be powered from a
supply that is not protected by a residual current detec-
tor (RCD).
Plug X1: NSG 3040 system interface IN – to be connected to NSG 3040 or to
another accessory, to X2 plug.
Power LED (green) shows if instrument is powered up
Error LED (red)
ERROR LED off:
No problem - accessory is ready to run
ERROR LED blinking:
Problem which may be solved by user intervention. Ex:
Interlock is activated - emergency button is pressed – overtemperature (for
MFO 6502).
ERROR LED on:
Problem which needs module repair – please contact your
nearest Teseq customer support center or sales representative.
Installation - connection to NSG 3040