133
Power LED (green) shows if instrument is powered up
Error LED (red)
ERROR LED off:
No problem - accessory is ready to run
ERROR LED blinking:
Problem able to be solved by user. Ex.: Interlock is
activated – emergency
button is pressed – overtemperature
ERROR LED on:
Problem which needs module repair – please contact your
nearest Teseq customer support center or sales representative.
Indicated max values reachable for environmental tem-
peratures below 30° C. For higher environment tem-
peratures internal temperature sensor might trip after
a few minutes.
Installation - connection to NSG 3040
1) Connect instrument power in to mains
2) Remove 25 way Sub D plug at rear of NSG 3040
3) Connect this connector to X2 of MFO 6502
4) Connect NSG 3040 25 way output to MFO 6502 X1 plug, using system
interface cable delivered with MFO 6502
5) Connect MFO 6502 to loop antenna INA 701, 702 or 703
6) In case of INA 702, verify that “Power“ plug is fitted to the coil interface
unit
7) Power on MFO 6502
8) Power on NSG 3040 main frame
Operation
The coil factor is given by the loop antenna manufacturer. For Teseq INA 701,
702 and INA 703 loop antennas this factor is labelled on the antenna and is also
indicated in the test report delivered with it.
The equipment should be switched off during installa-
tion and interconnection.