131
For proper operation of the plug and play detection
mechanisms it is strongly recommended to power on
first the MFO 6502 accessory and then the NSG 3040.
Powering on the NSG 3040 before the accessories
mayresult in a non detection of the accessories.
Power supply
+ 15 V, 0 V, -15 V
Sine wave
Signal generator
Power amplifier
Mains
input
Red
Black
50 Hz
60 Hz
Amplitude
control
Low/High
range
Controller
Plug X1 – System cable IN
25 Way Sub D - Male
Plug X2 – System cable OUT
25 Way Sub D - Female
Connection to NSG master
controller via system interface
cable
Termination with NSG Interlock
plug
Circuit diagram MFO 6502