48
NSG 3040 EMC test system
EUT OFF/EUT ON
Touch the “EUT Off/EUT On” button to switch EUT power off or on.
Note:
The EUT power function can work only in combination with an automated
accessory, such as a variac, step transformer or automated CDN.
RAMP VALUE
The “Ramp value” button is active only if a rampable parameter in the test
window is selected. All rampable parameters are identified by a small gray ramp
icon. This icon will turn red when a parameter is ramped.
Ramping mode
Touch the “Ramping mode” button (“Static” in the
example) to change the ramping mode from static to
linear. In linear mode the user can set Start, Stop and
Step values.
Start
Touch the “Start” button (“200 V” in the example). A
red frame is displayed around the field. Enter the Start
value using either the wheel or the keypad.
Stop
Touch the “Stop” button (“4800 V” in the example). A
red frame is displayed around the field. Enter the Stop
value using either the wheel or the keypad.
Step
Touch the “Step” button (“1 V” in the example). A red
frame is displayed around the field. Enter the Step
value using either the wheel or the keypad.
Ramping window for voltage parameter