Instruction Manual for TESA CLINOBEVEL 3
9
Function - 1 -
The key
is used to transmit measuring values to the
RS485 port. The measured values can also be transmitted via the
same interface to a PC or laptop for further processing (e.g.
hyperterminal).
Data format OUT port:
[sss xxxxxx sn.nnnnnn<cr>]
sss =
0 ... 255 - continuous number
x
xxxxxx =
Serial number and type of sensor
(example:
N2673L
TESA CLINOBEVEL 3)
sn.nn
sn.nnnnnn = +9.999999 - Positive out of measuring
range
-9.999999 - Negative out of measuring
range
other value - angular value in rad
e.g. +0.226349
transmission format:
asynchron, 7Bit, 2 Stopbits, No Parity, 9600 Baud
Function - 2 -
Cancel (unfreeze) the "
HOLD
"- function to return to the
measuring
mode
.
At the same time the HOLD value is sent to the RS-485 port.
Function - 3 -
Escape
function from the menu
ZERO/SELECT - key
Function - 1 -
The key
is used to
changing the scale in the display
increase / decrease the display range.
This function can, however, be disabled in the instrument settings.
Function - 2 -
The key
is used to select possible adjustments,
such as
menu selection
modification of a figure in the menu
HOLD - key
Function - 1 -
The key
serves for "freezing" a measuring value.
The measured value is displayed until the TESA CLINOBEVEL 3
returns to measuring mode by pressing the key
SEND
.
Function - 2 -
In the functions "
REL.ZERO
" and "
ABS.ZERO
" the key
is used for reading in the actual measuring value
again during the manual entering.
2.6 Batteries /
Rechargeable
batteries
The two batteries are shipped separately. It is strongly recommended to remove the batteries
during transport or longer storage.
The battery voltage is shown in the display e.g.
2
7
(2,7 Volt)
.