Instruction Manual for TESA CLINOBEVEL 3
23
7.7 Absolute
measurement
As a factory setting the TESA CLINOBEVEL 3 will be programmed for absolute measurement.
If this is not the case, select the function [Absolute]. After confirming this function with the key
the instrument is ready for measurements in the mode "ABSOLUTE".
The displayed measuring value corresponds to the value of the TESA CLINOBEVEL 3 minus
the ZERO-OFFSET.
Value displayed = Value of the TESA CLINOBEVEL 3 - “ZERO OFFSET”
7.8 Relative
measurement /
REL ZERO
Important preliminary remark:
The "REL ZERO OFFSET" determined for a relative measurement will be superposed to the
"ZERO OFFSET", e.g. determined by a reversal measurement.
The "REL ZERO OFFSET" will be stored in the TESA CLINOBEVEL 3 and can be re-called
again and again. When starting the next relative measurement, the REL ZERO OFFSET
entered or determined the last time will be displayed. The value can either be confirmed,
newly entered or set to zero.
VALUE displayed =
Value of the TESA CLINOBEVEL 3 - "ZERO-OFFSET" - "REL ZERO-OFFSET"
Abbreviated procedure with the key
:
Put the measuring instrument on the reference surface.
The display shows the value -7‘00‘‘. This corresponds to
the absolute inclination of the reference surface.
Set the TESA CLINOBEVEL 3 to the correct position and press the key
.
After 15 seconds the measuring value will automatically be
read.
During the measurement the display will graphically show
the current measurement.
Complete the measurement with the key
or with the remote control. After 15 seconds the measuring
value will automatically be read.
On the screen the display for the actual measurement will
appear now under consideration of the ZERO-OFFSET.